Medical MEB-1K5A24T AAA Product Change Notice


Delta reduces leakage current to increase user safety for MEB-1K5A24T AAA series medical power supplies. Please refer to the table below for information on the changes. Comprehensive testing and verification of the results have been conducted to ensure that all customers can expect consistent, high-quality product performance after these design changes.


If you require any further clarification, please refer to the latest specifications on the DeltaPSU website or reach out to us via our Contact Us Form.


Model List:


Model Name Description Old Revision
New Revision
Implementation Date
MEB-1K5A24T AAA Reduce leakage current 00 01 2025/10/16

Modification of MEB-1K5A24T AAA:


Specifications Before the Change Specifications After the Change
Touch Current (maximum) 0.1 mA @ Normal Condition
0.5 mA @ Single Fault Condition
0.1 mA @ Normal Condition
0.2 mA @ Single Fault Condition
Earth Leakage Current (maximum) 0.5 mA @ Normal Condition
1.0 mA @ Single Fault Condition
0.2 mA @ Normal Condition
0.4 mA @ Single Fault Condition

Related News

PMT2 Series Upgrade Notice

PMT2 Series Upgrade Notice PMT-xx2BA series upgrade according to new certification ...

Mehr

Product Change Notice: Update of DNV Certification Mark on Specification Label for CliQ M Series

The specification label for the CliQ M Series will be updated to reflect the revised DNV c ...

Mehr

Delta Product Label Trademark Update and the Removal of Representative Address

1. The Delta trademark on the spec label of each power supply unit will be updated. 2. ...

Mehr

Melden Sie sich jetzt für unseren Newsletter an

Seien Sie mit der/die erste um über neue Veranstaltungen, Neuigkeiten und Produkte informiert zu werden!

Melden Sie sich jetzt für unseren Newsletter an

Seien Sie mit der/die erste um über neue Veranstaltungen, Neuigkeiten und Produkte informiert zu werden!